Editorial Type:
Article Category: Research Article
 | 
Online Publication Date: 22 May 2025

The Performance of PEKK versus Zirconia Abutments for Screw-Retained Crowns in Two-Piece Zirconia Implants: an in vitro Study

DDS,
DDS,
PhD,
PhD,
DDS, and
DDS
DOI: 10.1563/aaid-joi-D-25-00014
Save
Download PDF

ABSTRACT

Objectives:

This study aimed to compare the in vitro performance of anterior and posterior crowns with screw-retained polyetherketoneketone (PEKK) or zirconia abutments on two-piece zirconia implants.

Materials and Methods:

Monolithic incisor and molar zirconia crowns (n=8 per group) were cemented on either PEKK or zirconia screw-retained abutments on zirconia implants. Molar and incisor crowns were also cemented on screw-retained titanium abutments on titanium implants as control. All specimens were subjected to combined mechanical loading (1.2 × 106 cycles of 50N, f = 1Hz) and thermal cycling (2 × 3000 × 5°C/55°C cycles of 2 min). Complications, such as loosening of the crowns and fractures, were observed. The surviving specimens were subjected to a fracture test. Descriptive statistics were applied to complications, survival times, and fracture forces.

Results:

The occurrence of complications depended significantly on location (anterior or posterior) and material (P<.0001). While the incisor test groups presented higher failure rates for both PEKK and zirconia abutments, the survival rates of the molar-shaped crowns of both test groups were comparable to those of the titanium control group. This highly significant interaction is caused by a highly significant difference in location for PEKK and zirconia (P<.05). The titanium control group showed uniform performance independent of the simulated site.

Conclusion:

PEKK abutments showed high survival rates for in vitro simulated molar sites. Screw-retained zirconia abutments have shown higher complication rates, especially in simulated anterior sites. Neither zirconia nor PEKK abutments seem to be recommendable in clinical application for anterior indication.

Contributor Notes

Corresponding Author Valentin Herber, Universitäres Zentrum für Zahnmedizin Basel (UZB) University of Basel Klinik für Oralchirurgie CH-4058 Basel, Switzerland Tel.: +41 61 267 25 33 Email address: valentin.herber@unibas.ch

Conflicts of interests: The authors declare no conflicts of interest regarding this manuscript.

  • Download PDF